Appendix A:
KnightStar 330
Setup
Checklist
Table 13. KnightStar 330 Setup Checklist
Procedure
Pass
Fail
General Exterior Appearance
Any dents, scratches, or loose parts that may indicate dropping or other abuse?
❑❑
Inlet baffle missing?
❑❑
Check the condition of A/C power cord.
❑❑
Check for fluid residue in and around KnightStar 330 openings and housing joints.
❑❑
Ensure that the inlet filter is clean and in place.
❑❑
Ensure that a new optional outlet air filter is used.
❑❑
KnightStar 330 Setup
Ensure that the KnightStar 330 is placed in such a manner that there is at least
one inch of clearance at the back of the device.
Connect one end of the A/C power cord into the rear panel of the KnightStar 330,
and the other end into an A/C wall outlet.
Turn the on/off button on. Both the yellow and indicators should flash for
approximately one second; the green indicator remains lit.
❑❑
CAUTION: If an error code appears on the display, or an alarm stays activated follow-
ing system power-up, turn the on/off button off. Then turn the on/off button on; if the
system fails again, the KnightStar 330 must be serviced before installation can con-
tinue.
Functional Test
With the unit turned on, select the CPAP mode and set an alarm volume of 0.
Set the CPAP prescription pressure to 3 cmH
2
O, then increase the pressure to
20 cmH
2
O. As the pressure increases, you should be able to hear the motor blower
speed increase.
❑❑
Set the alarm volume to 1.
Turn the KnightStar 330 off and wait for the motor to stop rotating.