Puritan Bennett KnightStar 330 User Manual

Browse online or download User Manual for Household fans Puritan Bennett KnightStar 330. Puritan Bennett KnightStar 330 User`s manual [en]

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - Ventilator

User’s ManualY-500009-00 Rev. GKnightStar ® 330Bi-Level® Ventilator

Page 2

6WARNINGS (continued)Be sure to clean the KnightStar 330 and the breathing circuit before use.The KnightStar 330 should never be used with improper or

Page 3 - Contents

7WARNINGS (continued)The KnightStar 330 and all other bi-level pressure devices should be used only with interfaces recommended by the device’s manufa

Page 4

8Note: At the end of the KnightStar 330's useful life, return the device to the manufacturer for proper disposal.WARNINGS (continued)However, the

Page 5 - Introduction

9SymbolsTable 1 lists descriptions for the various symbols that appear on the KnightStar 330. Table 1. SymbolsSymbol DescriptionAttention, consult acc

Page 6

10Symbol DescriptionAlarm conditionCE Mark: This device complies with the requirements of Medical Device Directive 93/42/EEC concerning medical device

Page 7 - Cautions and Warnings

11Symbol DescriptionUL mark, classified by Underwriters Laboratories Inc. with respect to electric shock, fire, and mechanical hazards only in accorda

Page 8 - WARNINGS:

12KnightStar 330 ComponentsThe KnightStar 330 components are shown in Figure 1.Figure 1. KnightStar 330 Components Power cordBlower outletTubing and p

Page 9

13System DescriptionControl Panel DisplayThe control panel display (shown in Figure 2) provides an easy-to-read format for mode, settings, and patient

Page 10

14Table 2: Control Panel ButtonsSymbol Name FunctionOn/Off Turns the KnightStar 330 on or off. Turn the KnightStar 330 on with a quick press and relea

Page 11

15▼/Delay/Ramp Button. Press the ▼/Delay/Ramp button to start the Delay/Ramp feature. When the delay is started, both inhale and exhale pressures will

Page 12

ii© Copyright 2000, 2003 Puritan-Bennett Corporation, 4280 Hacienda Drive, Pleasanton, CA 94588 U.S.A. All rights reserved.KnightStar® and Bi-Level® a

Page 13 - Symbol Description

16After the delay time has elapsed, pressure will slowly ramp up to the prescribed pressures.The Delay mode can be cancelled by again pressing the ▼/D

Page 14

17A high priority alarm is indicated by a flashing red LED, along with an audible alarm that beeps five times at intervals of ten seconds. A medium pr

Page 15

18KnightStar 330 Air Outlet AssemblyThe KnightStar 330 air outlet assembly (shown in Figure 4) consists of:• Air outlet • Optional outlet air filterAi

Page 16 - KnightStar 330 Components

19Setup and Use of the KnightStar 330NOTE: Before using the KnightStar 330 for the first time, read the section on “Cleaning” on page 25.Power On Self

Page 17 - System Description

203. Press the On/Off button located on the top panel of the KnightStar 330. The unit will go through a power-on self-test, and will then automaticall

Page 18 - Symbol Name Function

21Using the Delay/RampThe Delay/Ramp feature enables you to reduce the start pressure at the beginning of the night to make it more comfortable for yo

Page 19

222. If at any time you want to stop the delay, simply press the ▼/Delay/Ramp button again and the delay will be cancelled. You can also reset the del

Page 20 - Control Panel Indicators

23Using Supplemental OxygenIf your physician determines that you need supplemental oxygen, an oxygen source may be connected to the device.The oxygen

Page 21 - WARNING:

24Figure 5. Connecting the Oxygen Adapter to the Air OutletOxygen adapterOptional outlet air filterPort

Page 22 - Figure 4. Air Outlet

25CleaningIt is important to clean all of the equipment regularly. Do not use any cleaning method other than that described here.Cleaning the TubingBe

Page 23 - Power On Self-Test

iContentsIntroduction ...1Cautions and War

Page 24

26Cleaning the Inlet FilterInspect the inlet filter often by removing the inlet baffle (removable plastic cover on the back of the unit). Clean the fi

Page 25 - Using the Delay/Ramp

27Replacing the Optional Outlet FilterThe optional outlet filter is disposable, and should be inspected regularly and replaced when noticeably dirty o

Page 26

28Using the Optional HumidifierA humidifier may be used with the KnightStar 330 if the patient is experiencing nasal discomfort due to low moisture co

Page 27 - Using Supplemental Oxygen

29Figure 6. Placing the KnightStar 330 on Top of the Humidifier Housing2. Remove the reservoir from the housing and fill it to the FILL LINE (Figure 7

Page 28 - Optional outlet air filter

304. Slide the reservoir gently back into the housing.5. Connect the short humidifier tubing (Figure 8) between the KnightStar 330 and the inlet of th

Page 29 - Cleaning

31External BatteryWhen power from a wall outlet is unavailable, the KnightStar 330 can run on a 12 V external battery. Use a special cable from Purita

Page 30 - Cleaning the Inlet Filter

32Problem SolvingAny unusual system event results in one or all of the following: • Error code display • LED illumination• Audible alarm soundingTo mu

Page 31

33Table 3: Troubleshooting ChecklistProblem Indicators Possible Cause Corrective ActionNo airflow out of deviceNo alarm or displayed symbol.1. Interna

Page 32 - Using the Optional Humidifier

34Internal malfunctionAlarm and flashing LED. Displayed symbol is . The segment ## is the 2-digit error code.Internal electronic problem.Disconnect

Page 33 - Figure 7. Reservoir Fill Line

35Note: Consult your clinician to change the high and low pressure settings.Circuit leak Alarm and flashing LED. Displayed symbol is .Tubing circuit

Page 34 - Figure 8. Humidifier Tubing

iiProblem Solving ...32What the Patient and Caregiv

Page 35 - External Battery

36What the Patient and Caregiver Must KnowTable 4 lists topics that patients and caregivers must understand in order to use this device successfully.

Page 36 - Problem Solving

37.Table 4: Patient/Caregiver ChecklistThe patient and caregiver must understand:❑The need for bi-level ventilation.❑The schedule for ventilation.❑The

Page 37

38❑The ventilator alarm settings, and the purpose and function of each.❑How to respond to bi-level ventilator alarms.❑What to do if the bi-level venti

Page 38

39❑How to determine the quantity of oxygen being delivered, and how to adjust the quantity.❑Safety rules for the use of oxygen.❑How and why to monitor

Page 39

40❑How to recognize and respond to problems with the bi-level ventilator.❑How to recognize and respond to problems with the oxygen supply.❑Techniques

Page 40

41IndexAAir Outlet 18Filter 27AlarmsHigh Priority 17, 32Low Priority 17, 32Medium Priority 17, 32BBattery 31CCautions 3Cleaning 25Components 12Control

Page 41

42TTubing 25WWarnings 3

Page 43

Puritan-Bennett Corporation4280 Hacienda DrivePleasanton, CA 94588 USAToll Free: 1.800.6355267Authorized RepresentativeTyco Healthcare UK Limited154 F

Page 44

1IntroductionThe Puritan Bennett KnightStar 330 ventilator is intended to assist patients with ventilation when they are experiencing respiratory insu

Page 45

2The regimen for the treatment of your disease is prescribed by your physician. Your physician has prescribed various settings that are preset into th

Page 46 - Warnings 3

3Cautions and WarningsThe following words found in this document have special significance.WARNING: Means that there is a possibility of injury or dea

Page 47

4WARNINGS:In using this equipment, it is important that you read, understand, and follow the instructions and warnings in this user’s manual.This vent

Page 48 - Y-500009-00 Rev G

5WARNINGS (continued)To avoid electrical shock:• Do not use if power cord or plug is damaged• Unplug all power cords before cleaning• Do not use near

Comments to this Manuals

No comments