Puritan Bennett KnightStar 330 User Manual Page 38

  • Download
  • Add to my manuals
  • Print
  • Page
    / 48
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 37
34
Internal
malfunction
Alarm and flashing
LED. Displayed symbol
is .
The segment ## is the
2-digit error code.
Internal electronic
problem.
Disconnect power, then reapply power. If
condition persists, contact your home care
provider for repair.
Overpressure Alarm and flashing
LED. Displayed symbol
is .
Internal electronic
problem.
Disconnect power, then reapply power. If
condition persists, contact your home care
provider for repair.
High pressure Alarm and flashing
yellow LED. Displayed
symbol is .
Kinked or blocked
tubing.
Verify that the tubing has not collapsed,
and that there are no sharp bends.
Reposition the device, tubing, or
accessories, as applicable.
Low pressure Alarm and flashing
LED. Displayed symbol
is .
1. Tubing circuit leak,
or tubing is
disconnected.
2. Small, proximal
pressure tubing is
not connected to
port next to device
air outlet.
1. Reposition interface pillows or mask.
Check tubing connections at device air
outlet and patient interface. If tubing is
punctured or disconnected, replace it
or reconnect it, as applicable.
2. Verify proper tubing connection.
Disconnect tubing & reinstall, as appli-
cable.
Table 3: Troubleshooting Checklist (continued)
Problem Indicators Possible Cause Corrective Action
##
55