Troubleshooting
Trouble Probable Cause Remedy
Place cannula prongs in a clean glass
of water. If you observe bubbles
coming from your cannula, your unit
is working correctly.
Check each connection from the
unit to the cannula to ensure a tight
fit. If you are using a humidifier
bottle, check to see that it is screwed
on tight.
Adjust flow knob to point directly to
prescribed setting.
Call Apria Healthcare for
immediate delivery.
Remove cannula or mask and check
tubing for kinks or obstructions.
Replace with new cannula or mask
if needed.
Fill portable and call Apria
Healthcare location.
Wait 15 to 30 minutes for
connections to thaw.
No oxygen coming
from cannula or
mask
Portable unit cannot
be removed from
stationary system
Decreased awareness
of oxygen flow
Loose connections
Flow control knob is
not pointing directly
to setting
Unit is empty
Dirty or faulty cannula
or mask
Stationary unit
obstructed flow
Units are frozen
together