Introduction
1-6
KnightStar 330 User’s Manual Y-500009-00 Rev. J (6/06)
Always place the KnightStar 330 upright on a
firm, flat surface, and use only approved
accessories. Placing the device on uneven
surfaces, using it in an improper orientation,
or using unapproved accessories could result
in the device tipping over, causing damage or
possible patient injury.
The KnightStar 330 should be used with care
to avoid overheating the patient when the
room temperature exceeds 90
o
F (32.2
o
C),
since under certain conditions the patient
outlet gas flow can be as much as 6.7
o
F
(3.7
o
C) degrees warmer than room
temperature.
The KnightStar 330 equipment has been
tested and found to comply with the limits for
medical devices to IEC 601-1-2:2001 (or EN
60601-1-2:2001 or Medical Device Directive
93/42/EEC). This testing shows the device
provides reasonable protection against
harmful interference in a typical medical
installation. However, there is no guarantee
that interference will not occur in a particular
installation, for example, in the presence of
portable and mobile RF communication
devices. If this equipment does cause harmful
interference to other devices or is negatively
impacted by other devices, the user is
encouraged to try to correct the interference
by one or more of the following measures:
Reorient or relocate the devices
Increase the separation between the
devices
Connect the equipment to an outlet
on a different circuit
Consult the manufacturer or field
service technician for help