How to solve device problems
4
Y-500009-00 Rev. J (6/06) KnightStar 330 User’s Manual
4-3
High
pressure
Alarm and
flashing
yellow
LED.
Displayed
symbol is
.
Kinked or blocked
breathing circuit.
Verify that the breathing
circuit has not
collapsed, and that
there are no sharp
bends. Reposition the
device, breathing
circuit, or accessories, as
applicable.
Low pressure Alarm and
flashing
red LED.
Displayed
symbol is
.
1. Breathing cir-
cuit leaks or is
disconnected.
2. Small patient
pressure tube
is not con-
nected to port
next to device
air outlet.
1. Reposition nasal pil-
lows or mask. Check
breathing circuit
connections at
device outlet filter
and patient inter-
face. If breathing cir-
cuit is punctured or
disconnected,
replace it or recon-
nect it, as applica-
ble.
2. Verify proper breath-
ing circuit connec-
tion. Disconnect and
reconnect, as appli-
cable.
Interface/
mask leak
Alarm and
flashing
red LED.
Displayed
symbol is
.
Mask or breathing
circuit leaks or is
disconnected.
Reposition nasal pillows
or mask. Check
breathing circuit
connections at device
outlet filter and patient
interface. If breathing
circuit is punctured or
disconnected, replace it
or reconnect it, as
applicable.
Table 4-1: Troubleshooting checklist (continued)
Problem Indicators Possible Cause Corrective Action
P
P
L