Puritan Bennett KnightStar 330 User Manual Page 47

  • Download
  • Add to my manuals
  • Print
  • Page
    / 68
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 46
4
CHAPTER
4-1
Y-500009-00 Rev. J (6/06) KnightStar 330 User’s Manual
4
How to solve device problems
Any unusual system event results in one or all of the
following:
Error code display or alarm symbol
LED illumination
Audible alarm
Alarms are classified by priority:
HIGH priority—Indicated by a flashing RED LED
accompanied by an audible alarm signal (a series of
3 beeps, then 5 beeps, then 2 beeps at intervals of
approximately 6 seconds).
MEDIUM priority—Indicated by a flashing YELLOW
LED accompanied by an audible alarm signal (3 beeps
at intervals of approximately 25 seconds).
LOW priority—Indicated by a steadily lit YELLOW LED
on the control panel and no audible alarm.
In many cases, the alarm condition can be remedied by
patient or caregiver intervention.
To mute an alarm for one minute, press the
S/Alarm Silence key.
There is no audible alarm built into the
KnightStar 330 to indicate that the patient has
stopped breathing.
Page view 46
1 2 ... 42 43 44 45 46 47 48 49 50 51 52 ... 67 68

Comments to this Manuals

No comments